Quick review of electrode area / impedance within m8ta:
- {895} 500um^2
- {311} 490um^2 nominal; 900k
- {1040} 108um^2, plated from 5M to 1M.
- Neuronexus: 177, 413um, 700um, and 1250um.
- Suggest 177um for SUA, 413um for MUA.
- Community consensus seems to be that these electrodes don't last as long, though.
- Electroplating 177um^2 sites with PEDOT:PSS reduces impedance to 23k {1388}
- {823} 122um^2 nominal
- {736} 500um^2
- {1027} (Utah) 1600um^2
- Impedance: ~ 220K +-91K (in vivo -- large variance)
- Blackrock site lists impedance @ 400k
- SIrOF Utah array -- 3100um^2 (3.1e-5cm^2) -- large!
- Impedance ~50K according to [www.blackrockmicro.com/userfiles/file/Microelectrode%20Arrays.pdf Blackrock product brochure].
- PMID-20124668 (Utah again) 2000um^2, 125k Pt, 6k SIROF.
- Neuropixel: 144um^2 acid-etched TiN
- Carbon fiber: ~38 um^2, PEDOT:PSS or PEDOT:pTS started ~ 4M, plated down to ~ 130k initial, went up to 2M pSS, 840k pTS.
|